Compact spectrometers are widely used in educational applications, in field work where simplicity and compactness are crucial, and in research laboratories when a quick but accurate answer is needed. They are easily transportable, have no moving parts, and offer good spectral resolution. They are also very versatile, since they integrate easily into experimental setups with different light sources, cuvettes and other sample holders, optical fibers, lenses, and many other elements.
Axiom Optics’ compact spectrometer range now runs from the UV/Vis all the way into the mid-infrared. GoyaLab’s IndiGo instruments cover the visible and near-IR, the d-Vision from Sphere Photonics covers the visible through the short-wave infrared, and NLIR’s MIDWAVE mid-IR spectrometer covers 2.0 to 5.0 µm, where molecules have their strongest absorption bands.
From the visible to the mid-infrared
Pick a spectral region to highlight the instruments that cover it.
Visible light drives electronic transitions, so visible spectra reveal color, dye and pigment concentration, filter transmission, fluorescence, and the emission lines of light sources. This is the home ground of the GoyaLab spectrometers described below.
Near-infrared absorption comes from overtones and combinations of molecular vibrations. These bands are weak, so NIR light penetrates deep into a sample, which suits bulk analysis of food, grain, and pharmaceuticals, often straight through glass or plastic packaging.
720 to 1050 nm, 2 nm (FWHM) resolution, in the same compact handheld format as the IndiGo
Moving into the short-wave infrared reaches the stronger first and second overtones, including the water band near 1450 nm. VIS-SWIR spectra are used for moisture measurement, plastics sorting, and identifying materials by their spectral signatures.
500 to 1700 nm on a single InGaAs CMOS sensor, 0.7 to 3.1 nm resolution, 10 µs to 1 s integration, 16-bit, USB-C (by Sphere Photonics)
Between 2 and 5 µm sit the fundamental stretching vibrations of O-H, N-H, and C-H bonds and the strongest absorption bands of gases such as CO2, CO, CH4, and N2O. Fundamental bands absorb far more strongly than the NIR overtones, which makes the mid-IR the region of choice for gas sensing, combustion diagnostics, and fast chemical monitoring.
2.0 to 5.0 µm upconversion spectrometer by NLIR: no cryogenic cooling, no moving parts, fiber-coupled
2.2 to 5.0 µm center wavelengths, up to 10 GHz electrical bandwidth
Converts 2.0 to 5.3 µm light to 695 to 886 nm for silicon and GaAs detectors
Broadband mid-IR sources designed as companions to the MIDWAVE
Most spectrometers on the market are based on a Czerny-Turner optical configuration, where light enters through a slit and hits a concave mirror that reflects and collimates it onto a diffraction grating. The diffracted light is then reflected and focused by a second concave mirror onto a detector. GoyaLab has developed and patented an alternative optical configuration based on axial transmission. A transmission configuration, instead of the folded reflective design required in Czerny-Turner instruments, enables significantly higher light throughput and therefore higher sensitivity.

GoyaLab spectrometers combine this axial transmission configuration, which puts more light on the detector, with a 2D CMOS detector. Vertical binning on a 2D detector inherently gives higher signal-to-noise levels than a 1D detector array.
Mid-infrared light is hard to detect directly. Conventional mid-IR detectors are slow, noisy at room temperature, or need cryogenic cooling. NLIR solves this with upconversion: inside the spectrometer, the incoming mid-IR light mixes with a high-power pump laser in a nonlinear crystal, and through sum-frequency generation each mid-IR photon is converted into a near-visible photon. Fast, low-noise silicon detectors, the same kind used in visible spectrometers, then read the converted spectrum.
Schematic for illustration. The pump wavelength shown is consistent with the Wavelength Conversion Module’s published mapping of 2.0 to 5.3 µm input to 695 to 886 nm output.
According to NLIR, the result is a mid-IR spectrometer with no cryogenic cooling and no moving parts, fiber-coupled and ready for both laboratory work and rugged industrial process monitoring. The same upconversion principle powers NLIR’s single-wavelength detectors and the stand-alone Wavelength Conversion Module.
| Czerny-Turner | GoyaLab axial transmission | NLIR upconversion | |
|---|---|---|---|
| Light path | Folded, two concave mirrors | Straight-through transmission | Mid-IR mixed with a pump in a nonlinear crystal |
| Detector | Typically 1D array | 2D CMOS with vertical binning | Silicon, after conversion to near-visible |
| Main benefit | Established, flexible design | Higher throughput and signal-to-noise | Fast, sensitive mid-IR without cryocooling |
| Range at Axiom | n/a | 380 to 1050 nm (IndiGo, IndiGo NIR) | 2.0 to 5.3 µm (MIDWAVE and modules) |
| Instrument | Range | Resolution | Best for |
|---|---|---|---|
| GoSpectro | Visible | Better than 10 nm | Teaching and smartphone field checks |
| IndiGo | 380 to 780 nm | Under 1.5 nm | Absorbance, filters, light sources, fluorescence |
| IndiGo NIR | 720 to 1050 nm | 2 nm | Near-IR overtone analysis |
| d-Vision | 500 to 1700 nm | 0.7 to 3.1 nm | Moisture, plastics, material ID |
| MIDWAVE | 2.0 to 5.0 µm | 6 cm⁻¹ | Gas sensing, combustion, fast process monitoring |
The IndiGo spectrometer comes with PC software, SpectroLab, for controlling the spectrometer, acquiring data in real time, and analyzing the acquired spectra. SpectroLab is an intuitive and robust platform that also detects and controls the accessories attached to the IndiGo. Its features include:
SpectroLab also saves spectra in formats that work with other data analysis tools, including .csv, .txt, and the widely used spectroscopy .spc format.
The MIDWAVE runs on NLIR’s NLIzeR software, with APIs for MATLAB, Python, and C for integrating mid-IR spectra into automated test stations and process control loops.

A modular spectrometer is the ideal tool for detailed characterization of the transmission properties of spectral filters, or for identifying mislabeled or misplaced optical elements in a laboratory. In this example, the detailed transmittance of a bandpass filter with a central wavelength of 500 nm was measured with the IndiGo spectrometer. The setup is simple, requiring only the IndiGo spectrometer and a broadband light source.
Filter: Thorlabs FB500-40, CWL = 500 ± 8 nm, FWHM = 40 ± 8 nm, minimum transmission at CWL = 70%

The resulting transmission spectrum was compared with the filter’s Thorlabs spec sheet and with a second measurement made with an Ocean Insight Flame spectrometer and the same light source. Filter manufacturing involves a complex vacuum layer deposition process that produces normal variations in overall transmittance, central wavelength, and bandwidth, so detailed transmission characterization is essential when filters are used for specific applications. Both the IndiGo and Flame measurements, each a single acquisition (no averaging) with the same light source, reveal the expected differences from the spec sheet. Notably, the 2D detector in the IndiGo gives lower noise thanks to vertical averaging over many pixels.

Neutral density filters attenuate light by a nominal fraction over a wide wavelength range. Their nominal optical density (OD) indicates the magnitude of the attenuation. Percent transmission (%T) and OD are related by:
%T = 100 × 10−OD
For example, a neutral density filter with an OD of 1.0 transmits 10% of the light, and filters with an OD above 2.0 transmit well under 1%. A simple way to characterize the full transmission profile of a neutral density filter is to use a spectrometer and a broadband light source. The spectrometer needs enough dynamic range to measure the unattenuated source and enough sensitivity to quantify the light the filter transmits.
Optical density to transmission
A demonstration measurement on an OD 3.0 filter was performed with the GoLight broadband light source and the IndiGo spectrometer. In a single-shot measurement with a 250 ms integration time, the IndiGo correctly quantified a transmission of 0.1% without noise significantly affecting the result across most of its wavelength range, demonstrating the spectrometer’s sensitivity over its entire 1000:1 dynamic range.

Using the averaging feature, transmission levels of 0.01% and below near 400 nm could be quantified for the OD 3.0 filter:

The same transmission method extends into the mid-infrared. With the MIDWAVE and a thermal infrared light source, mid-IR bandpass filters, such as the 4.26 µm filters used in NDIR CO2 sensors, and IR windows can be characterized across 2.0 to 5.0 µm.

Equipped with the chemical analysis module, the IndiGo spectrometer performs spectrophotometric analysis by visible light absorption to determine the concentration of a solution using the Beer-Lambert law. According to the Beer-Lambert law, the absorbance of a solution at a given wavelength, A(λ), depends on the solution’s concentration c, the path length of the light through the sample l, and the extinction coefficient of the absorbing species at that wavelength, ε(λ). Within a certain concentration range the relationship between absorbance and concentration is linear, and the slope obtained from a series of dilutions of known concentration can be used to determine the unknown concentration of a solution of the same species by measuring its absorbance:
A(λ) = ε(λ) l c
In this application, a blue food colorant was diluted and used as a stock solution to prepare a series of serial dilutions for a standard curve of absorbance vs. concentration.

Each serial dilution halved the concentration, and its absorption spectrum was measured in a single shot with a 65 ms integration time. 1.0X, 2.0X, and so on refer to the dilution factor from the stock solution.

From the absorption spectra of the serial dilutions, a standard curve of absorbance vs. concentration (here, dilution factor) can be built at any wavelength. The graph below shows standard curves at three wavelengths. Strong absorbance gives more sensitive measurements, as seen in the standard curve near the absorption maximum at 630 nm, although other wavelengths can be used to avoid spectroscopic interferences at the expense of sensitivity. Over the concentration range of this experiment, all standard curves show excellent linearity, which demonstrates the IndiGo’s potential for quantitative concentration measurements.

The Beer-Lambert law is also the basis of mid-IR gas analysis. Because fundamental vibrational bands absorb far more strongly than visible or near-IR features, a mid-IR measurement reaches much lower concentrations for the same path length, and the MIDWAVE’s high-speed readout makes it possible to follow concentration changes in real time.
In this application, the resolution and sensitivity of the IndiGo spectrometer were pushed to the limit by measuring the visible absorption spectrum of iodine vapor in a cuvette with a standard 10 mm absorption path. The IndiGo was coupled with the chemical module, which includes a high-intensity flat-spectrum white light source for absorption measurements in the visible range. A small amount of solid iodine was sealed in the cuvette and heated with a heat gun to increase the iodine vapor concentration by sublimation.

Excitation of diatomic iodine gas into its first bound electronic excited state occurs above 500 nm, and the distinct absorption features correspond to changes in vibrational quantum number. The spacing between these features reveals the separation of the vibrational energy levels of the excited electronic state. The iodine absorption spectrum was obtained in a single-shot acquisition with a 65 ms integration time. For comparison, it is shown next to the iodine spectrum from a high-resolution Agilent Cary 100 UV-Vis spectrophotometer with a 0.278 m Czerny-Turner monochromator. The Cary 100 is equipped with a 1200 lines/mm grating, a tungsten-halogen/deuterium light source, and an R928 photomultiplier tube detector, and has a wavelength accuracy better than 0.04 nm and a resolution under 0.2 nm.

The overall agreement between the IndiGo and Cary 100 spectra is very good. Some differences stand out, mainly on the left side of the spectrum. The Cary 100 has a heating block that held the sample at a constant 75 °C, while the IndiGo spectrum was taken after heating the cuvette outside the spectrometer, so absorbance values are not expected to match quantitatively. Also, the rovibronic transitions on the left of the spectrum get progressively closer together (they correspond to higher vibrational levels in the excited state potential) and need higher resolution to be resolved. Within the resolution limits of the IndiGo, the agreement is very good, and the IndiGo shows enough sensitivity for a gas-phase absorption measurement even in a single-shot acquisition.
In the visible, iodine’s vibrational levels show up only as structure on an electronic transition. To measure molecular vibrations directly, move to the mid-infrared, where the MIDWAVE records the fundamental vibrational bands of gases and functional groups, as shown in the next section.
The iodine example probes electronic transitions, with vibrational structure riding on top. In the mid-infrared, molecules absorb directly through their fundamental vibrations, without any electronic excitation. Each functional group and gas has its own band positions, which makes 2 to 5 µm spectra a direct way to identify and quantify species such as CO2, CO, methane, and hydrocarbons.
Mid-IR band finder: what can you measure between 2 and 5 µm?
Moisture and humidity, water in solvents and process streams, and the O-H stretch of alcohols. Water vapor is also a key combustion product.
N-H stretching of amines, amides, and ammonia (NH3), used in emissions control and chemical process monitoring.
Natural gas leak detection, biogas and landfill monitoring, and greenhouse gas emissions.
C-H stretching in polymers, fuels, oils, and solvents: material identification and hydrocarbon process control.
Combustion monitoring, fermentation, breath analysis, and the reference band of NDIR CO2 sensors.
Emissions and agricultural soil flux monitoring, plus anesthetic gas measurement.
Combustion efficiency and safety monitoring in burners, engines, and furnaces.
Approximate band positions of the strongest fundamental absorptions. Upconverted wavelengths assume a pump near 1064 nm, consistent with the Wavelength Conversion Module’s published 2.0 to 5.3 µm input and 695 to 886 nm output ranges.
NLIR’s upconversion instruments are used for:
Read more in Axiom’s announcement: Axiom Optics adds the fastest and most sensitive mid-infrared spectrometers and detectors to its portfolio.
What do you need to measure?
Visible absorption, Beer-Lambert concentration curves, filter transmission, and light source emission from 380 to 780 nm, with SpectroLab software on a PC or the mobile app.
Turns a smartphone or tablet into a handheld spectrometer for teaching, demonstrations, and fast field checks across the visible range.
720 to 1050 nm with 2 nm resolution for overtone-band analysis of bulk samples in the same compact format.
500 to 1700 nm on one InGaAs CMOS sensor, covering the water band near 1450 nm and the overtones used for plastics and material identification.
Full 2.0 to 5.0 µm spectra of CO2, CO, CH4, N2O, and water vapor, fast enough for combustion and process dynamics, with no cryogenic cooling. Pair it with a thermal infrared light source for transmission measurements.
Narrowband mid-IR detection with up to 10 GHz electrical bandwidth and minimum detectable power from 1 fW to 1 nW, for laser pulse analysis and the fastest or weakest signals.
Converts 2.0 to 5.3 µm light into 695 to 886 nm so an existing silicon or GaAs spectrometer or camera can read it, at 2.5 cm⁻¹ output spectral resolution.
Not sure which spectral range fits your application? Axiom Optics applications engineers can help you choose between visible, NIR, SWIR, and mid-IR instruments and configure the right light source and sampling accessories.
Contact Axiom OpticsCompact spectrometers measure how light is emitted, absorbed, transmitted, or reflected across a range of wavelengths in an instrument small enough to carry. They are used in teaching labs, field work, and research labs for tasks such as filter characterization, concentration measurements with the Beer-Lambert law, light source testing, and molecular spectroscopy.
The IndiGo covers 380 to 780 nm, the IndiGo NIR 720 to 1050 nm, the d-Vision 500 to 1700 nm, and the MIDWAVE mid-IR spectrometer 2.0 to 5.0 µm, so together they span the visible through the mid-infrared.
The incoming mid-infrared light is mixed with a strong pump laser in a nonlinear crystal. Through sum-frequency generation, each mid-IR photon combines with a pump photon to create a near-visible photon, which fast, low-noise silicon detectors then measure. This removes the need for cryogenically cooled infrared detectors.
An FTIR uses a scanning interferometer with moving mirrors, and a scan typically takes seconds. It offers broad spectral coverage and high resolution for detailed chemical analysis. The NLIR MIDWAVE has no moving parts and needs no cryogenic cooling, and it reads out complete 2.0 to 5.0 µm spectra at high speed, which suits fast process monitoring, combustion, and transient measurements. The two are complementary.
The 2 to 5 µm range contains the fundamental stretching bands of O-H, N-H, and C-H bonds and strong absorption bands of gases including water vapor, methane, carbon dioxide (4.26 µm), nitrous oxide, and carbon monoxide.
Yes. In a single 250 ms acquisition, the IndiGo correctly quantified the 0.1% transmission of an OD 3.0 filter across most of its wavelength range, and with averaging it quantified transmission of 0.01% and below near 400 nm.